Can a Smartwatch or Ring Really Measure Blood Pressure?
Some wearables inflate a tiny cuff, some estimate from your pulse after you calibrate with a real cuff, and some only flag a pattern. They are not the same measurement.
Evidence rating for the proposition “Cuff-calibrated cuffless devices track blood pressure changes away from their calibration value”: Uncertain · KnowYourPrime Evidence Registry, Finding KYP-ER-0022
Only one kind can measure it the way a home monitor does. Watches with a small inflatable cuff in the strap, the Huawei Watch D line and Omron's HeartGuide, have passed the international cuff standard at rest, though independent tests outside the lab found larger errors. Everything else uses an optical pulse sensor, and that comes in two forms. Samsung and Whoop show numbers and the Hilo band shows trends, but only after you calibrate them with a real cuff, and independent studies find their readings stay pulled toward that calibration value, drift at night, and can miss a real change such as a fall after a medication increase. Apple, Oura, Google and RingConn show no numbers at all and look for a pattern over about a month. Apple's version is FDA-cleared, and in its validation study of 1,863 people it caught 41.2 percent of those with hypertension while rarely alerting people without it. The American Heart Association and the European hypertension societies say none of these optical features should be used to diagnose or manage high blood pressure yet.
What a watch or ring is actually measuring
They differ in what they physically measure. An inflatable cuff squeezes the artery and reads the pressure, the same principle as an upper-arm home monitor. An optical sensor reads the shape and timing of the pulse, which is related to blood pressure through the stiffness and tone of the blood vessels but is not pressure itself. To turn that into millimeters of mercury, a device needs either your own cuff readings as an anchor or a general model that has to hold for everyone who wears it.
So the products sort into the groups below, and a number from one group and an alert from another cannot be compared directly.
| Approach | Shows a number? | Needs your cuff? | Examples (October 2026) |
|---|---|---|---|
| Inflatable cuff in the strap | Yes | No | Huawei Watch D, D2; Omron HeartGuide |
| Optical estimate calibrated to your cuff | Yes | Yes | Samsung Galaxy Watch, Whoop, Hilo band |
| Optical estimate with no calibration | Yes | No | Happy Ring (FDA-cleared, September 2026) |
| Optical pattern over about 30 days | No: an alert or category | No | Apple Watch, Oura, Google Pixel Watch, RingConn |
Grouping by measurement method, from each company's own documentation read in October 2026: Samsung press release (March 2026), Whoop accuracy page (June 2026), Apple validation paper (September 2025), Oura support (October 2026), Google support (2026), RingConn (2026).
Why calibrated readings drift away from the truth
A calibrated optical estimate knows your pressure at the moment you calibrated. Between calibrations it has to infer any change from the pulse waveform, and that is where independent studies find the weakness. Samsung asks for cuff calibration every 28 days, and Whoop for three cuff readings at setup and one every 30 days after. Samsung's own validation and the studies of each Galaxy Watch generation are covered in how accurate Samsung and Whoop blood pressure is.
On a Galaxy Watch Active 2 worn alongside 24-hour ambulatory monitoring in 40 people, readings showed a bias toward the calibration point, overestimating low pressures and underestimating high ones (Frontiers in Cardiovascular Medicine, 2022). Over 28 days on a Galaxy Watch 5 in 37 people, average agreement with a cuff was close, but once true pressure was 10 points away from the calibration value the average error rose to 3.4 systolic and 5.1 diastolic, and the authors concluded accuracy declined as values diverged from the calibration point (Digital Health, 2026).
The Hilo band (formerly Aktiia) agreed with 24-hour monitoring during the day but read night-time systolic pressure 15.5 points higher and understated the overnight dip in 41 people. In three patients whose medication was increased, the home cuff recorded an average fall of 19.7 systolic and 11.5 diastolic, and the band recorded 1.0 and 0.8 (Journal of Hypertension, 2023). A pulse-transit device in 166 people showed the same pattern: the further pressure moved from the calibration reading, the larger the gap between it and the cuff (Journal of Hypertension, 2024).
Across studies of wearable cuffless devices against 24-hour monitoring, daytime averages agreed within acceptable limits but night-time readings were off by 4.48 systolic and 5.64 diastolic on average, and the authors concluded the devices are not yet reliable for night-time blood pressure (American Journal of Preventive Cardiology, 2025).
A systematic review of 245 studies of wearable blood pressure, most using pulse transit time or pulse wave velocity, found accuracy moderate to high for diastolic pressure but more variable for systolic, with calibration still needed and external validation rare (Archives of Gynecology and Obstetrics, 2026).
In 67 people wearing a watch-like optical sensor on the arm or wrist alongside a standard 24-hour cuff monitor, 24-hour average systolic readings differed by 1.8 mm Hg on average, but individual differences ran from 13.9 below to 10.3 above (Scientific Reports, 2023).
4 studies
- Samsung Galaxy Watch Active 2 vs 24-h ambulatory monitoring in 40 people: proportional bias toward the calibration point (low BP overestimated, high BP underestimated); sensitivity 83% and specificity 41% for hypertension.
- Samsung Galaxy Watch 5 vs Omron M4 over 28 days in 37 people: mean differences -0.34/0.62 mmHg with little drift, but 3.4/5.1 mmHg when reference BP was 10 mmHg from the calibration point.
- Aktiia band in 41 people: night-time systolic 15.5 mmHg above ambulatory monitoring and nocturnal dip understated by 14.2 mmHg; after medication up-titration in 3 patients, -1.0/-0.8 mmHg on the band vs -19.7/-11.5 mmHg on home cuff (P = 0.03).
- In 166 people, the cuff's 24-h mean differed from the calibration reading by 7.4 mmHg systolic on average, the cuffless device's by 1.8; the larger the change after calibration, the larger the gap between devices.
Watches with a real cuff in the strap
The Huawei Watch D line hides a small air bladder in the strap that inflates around the wrist, so it measures pressure directly and needs no calibration. Omron's HeartGuide, cleared by the FDA in 2018, works the same way. In 85 adults the original Watch D averaged 1.4 systolic and 0.2 diastolic below the reference, with spreads of 6.47 and 5.85, and met the international cuff standard (Blood Pressure Monitoring, 2022). The Watch D2 met it again in 85 adults at rest, averaging 1.6 systolic and 1.9 diastolic below the reference with spreads of 6.4 and 5.0, and agreed with 24-hour ambulatory monitoring in 35 adults, including at night (Hypertension Research, 2025).
The standard allows an average error of up to 5 with a spread of up to 8, so a watch can pass it and still be several points off on any one reading. A wrist reading also depends on where the wrist sits relative to the heart, which the European cuffless protocol tests separately.
Outside those validations the results are less clean. In 41 outpatients wearing a Watch D2 and a validated 24-hour monitor at the same time, the watch read systolic pressure 6.8 points low on average, and only 47 percent of systolic readings landed within 10 points of the reference, below the standard's acceptability threshold (Revista Clínica Española, 2026). In 100 patients from an ordinary clinic, Omron's HeartGuide, another watch with a cuff in the strap, read systolic pressure about 8.5 points low and did not pass all of the standard's criteria (Clinical Research in Cardiology, 2024). The European Society of Hypertension's 2026 consensus statement on these watches judges the static validations generally acceptable and the evidence for 24-hour use limited (Journal of Hypertension, 2026).
A ring cleared without cuff calibration
In September 2026 the FDA cleared Happy Health's blood pressure software for its Happy Ring, which reads light, skin conductance, motion and temperature from a ring on the index finger and reports systolic and diastolic numbers with no cuff calibration and no age or sex input. The clearance covers spot checks in adults aged 22 to 50 at home or in a clinic. It excludes people with arrhythmias such as atrial fibrillation, severe vascular disease or Raynaud's, people taking vasoactive drugs, pregnancy, and readings taken during motion.
The accuracy evidence is the company's own study as summarized by the FDA: 85 people compared against an arterial line, meeting both criteria of the cuff standard. No independent study has appeared in PubMed. A resting comparison also cannot show whether readings follow a change in pressure, the weakness the European cuffless tests were written to catch.
Alerts without numbers: Apple, Oura and Google
Apple Watch Series 9 and later and Ultra 2 and later can send a hypertension notification. Apple says the feature does not measure blood pressure or require calibration; it analyses 60-second optical samples taken about every two hours while you sit still during the day, over 30-day windows, and is for adults 22 and over with no prior diagnosis, not during pregnancy. The FDA cleared it in September 2025. In Apple's study of 1,863 people, it alerted 41.2 percent of those with hypertension, 29.6 percent at stage 1 and 53.7 percent at stage 2, and did not alert 92.3 percent of those without it.
An alert and a silent watch carry very different weight. Take a group where 40 in 100 people have hypertension they do not know about. Running Apple's own figures, about 78 in 100 of those alerted would have it, and about 30 in 100 of those who got no alert would still have it.
Oura added Blood Pressure Signals and Nighttime BP for Gen3 and later rings with a membership. Signals looks at 30 days of nighttime pulse data, needs at least 15 nights, and only notifies when it finds a pattern; Nighttime BP sorts the overnight dip into typical, reduced, pronounced or rising, without numbers, and the Oura blood pressure guide covers what each category has meant in cuff research. Oura's own blog reports 62 percent sensitivity and 91 percent specificity for hypertension, and for dipping 84 percent and 69 percent in 134 people against 48-hour monitoring; these figures have not been published in a journal. Google's blood pressure trends on Pixel Watch 3 and later show High, Not high or Not enough data, and Google states the feature is not a prescreener for hypertension. RingConn Gen 3 shows trends and states it is not a medical device.
What counts as validated, and what the medical societies say
The cuff standard tests whether readings agree on average and in spread. For cuffless devices the European Society of Hypertension recommends up to six tests, including whether readings follow a fall in pressure after treatment, a rise during exercise, the change from awake to asleep, and whether calibration holds over time (Journal of Hypertension, 2023). The FDA's January 2026 draft guidance for cuffless devices asks for a change test in which pressure moves at least 15 points systolic or 10 diastolic away from the calibration value. Whoop's published figures and Samsung's announcement do not report one.
The American Heart Association's scientific statement concludes that the devices on the market have not yet been adequately vetted for accuracy and efficacy (Hypertension, 2026). The European Society of Hypertension notes that its guidelines do not recommend cuffless devices for the diagnosis and management of hypertension (Journal of Hypertension, 2022), and a 2026 European scientific statement does not recommend their use in clinical decisions (European Journal of Preventive Cardiology, 2026).
Why most launched without FDA clearance: the FDA's January 2026 wellness guidance allows an optical blood pressure value as a general wellness output if it is validated and the product makes no diagnostic claims. Samsung, Whoop, Oura, Google and RingConn all label their features as wellness or not for diagnosis. Apple's notification, Happy Health's ring software and Aktiia's G0 system, the clinical version of the Hilo band, went through FDA 510(k) clearance. The Hilo Core sold to US consumers since July 2026 is labeled general wellness, and Hilo states the clearance does not apply to it.
The strongest validation result so far is for a finger device, not a watch. A finger-worn optical monitor met all seven European Society of Hypertension validation scenarios in 768 people, including recalibration, exercise and asleep tests, with mean differences of 5 mm Hg or less (Journal of the American Heart Association, 2026).
What this does not settle
Much of the validation is short, at rest, and often funded by or linked to the manufacturer. Few studies test the same device over months, during sleep, or after a real change such as starting medication, which is when a home reader most wants the number to move. Results for one model do not carry over to the next.
Common questions
Does the Apple Watch measure blood pressure?
No. Apple Watch Series 9 and later and Ultra 2 and later can send an FDA-cleared hypertension notification after 30 days, but Apple states the feature does not measure blood pressure or show a reading. In Apple's study it alerted 41.2 percent of people with hypertension.
Does the Samsung Galaxy Watch measure blood pressure accurately?
It shows systolic and diastolic estimates on Galaxy Watch4 and later after calibration with an upper-arm cuff every 28 days, and Samsung says it is not intended to diagnose high blood pressure. Independent studies find good average agreement near the calibration value and growing error as true pressure moves away from it.
Do smart rings measure blood pressure?
Not as a reading. Oura's Blood Pressure Signals and Nighttime BP and RingConn's Vascular Health Insights show patterns, categories or trends, without systolic or diastolic numbers. Oura's accuracy figures are its own and have not been published in a journal.
Is there a blood pressure ring that does not need a cuff?
Happy Health's Happy Ring is the only one with FDA clearance for that, granted in September 2026 for spot checks in adults aged 22 to 50. Its accuracy data comes from the company's own study of 85 people against an arterial line, with no independent study published yet.
Which smartwatches have passed the blood pressure standard?
In published studies, watches with a built-in inflatable cuff: the Huawei Watch D and Watch D2 met the AAMI/ESH/ISO universal standard at rest, and the European Society of Hypertension also counts Omron's HeartGuide among the validated watches, while independent tests outside the lab have found larger errors. Among the consumer smartwatches and rings named here, none with an optical-only method has published results covering the European Society of Hypertension's full set of cuffless tests.
Does Whoop measure blood pressure?
Whoop's Blood Pressure Insights gives a morning range estimated from overnight data after three cuff readings, with a new cuff reading every 30 days. Whoop reports an average error of 5.03 systolic and 3.64 diastolic against a cuff in its own data, and says the feature is not a medical device.
Can a smartwatch replace a blood pressure cuff?
Not according to the American Heart Association and the European hypertension societies, which do not recommend cuffless devices for diagnosing or managing hypertension. The calibrated optical features themselves depend on a cuff for their readings. What counts as normal at each age is covered in blood pressure by age.
KnowYourPrime Evidence Registry: Finding KYP-ER-0020: Accuracy of Smartwatches With a Built-In Blood Pressure Cuff · Finding KYP-ER-0021: Cuffless Blood Pressure Accuracy at Night · Finding KYP-ER-0022: Do Calibrated Cuffless Blood Pressure Readings Track Real Changes?
Sources
- Hypertension Notification Feature on Apple Watch (validation paper)
- 510(k) K250507, Hypertension Notification Feature (HTNF), Apple Inc.
- 510(k) K250415, G0 Blood Pressure Monitoring System, Aktiia SA
- Samsung's Blood Pressure Monitoring Feature Now Available to U.S. Users in Samsung Health
- How Accurate is WHOOP Blood Pressure Insights?
- Heart Health Hub (Blood Pressure Signals, Nighttime BP)
- What PPG Can Reveal: New Cardiovascular and Respiratory Research Findings
- Blood pressure trends (Google Health)
- RingConn Gen 3: Release Date, Blood Pressure Insights and Alerts
- Smartwatch-Based Blood Pressure Measurement Demonstrates Insufficient Accuracy
- Long-term accuracy and stability of blood pressure measurements from a smartwatch: Prospective validation study
- Evaluation of the ability of a commercially available cuffless wearable device to track blood pressure changes [conflict statement: Aktiia supplied the devices tested, and two authors have been paid by Aktiia]
- Ability of a 24-h ambulatory cuffless blood pressure monitoring device to track blood pressure changes in clinical practice
- Comparing the accuracy of continuous blood pressure monitoring using wearable cuffless devices with conventional 24-hour ambulatory blood pressure monitoring: A systematic review and meta-analysis
- Validation of the watch-type HUAWEI WATCH D oscillometric wrist blood pressure monitor in adult Chinese
- Accuracy of a wearable watch-type oscillometric blood pressure monitor in rest and ambulatory blood pressure measurements
- European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices
- Cuffless blood pressure measuring devices: review and statement by the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability
- Cuffless Devices for the Measurement of Blood Pressure: A Scientific Statement From the American Heart Association
- Cuffless Blood Pressure Monitoring Devices: Technical Foundations and Clinical Implications
- Cuffless Non-invasive Blood Pressure Measuring Devices: Clinical Performance Testing and Evaluation (draft guidance)
- General Wellness: Policy for Low Risk Devices (guidance)
- Clinical reliability of a wrist-worn oscillometric ambulatory blood pressure monitor: an independent agreement evaluation
- Performance of wearable watch-type home blood pressure measurement devices in a real-world clinical sample
- Wearable wrist-watch type cuff oscillometric blood pressure monitors: consensus statement by the European Society of Hypertension Working Group on blood pressure monitoring
- 510(k) K254000, Happy Health Blood Pressure Monitor (summary)
- 510(k) K182481, Omron HEM-6410T-ZM wrist blood pressure monitor (HeartGuide)
- Is Hilo Core FDA cleared?
- Non-invasive blood pressure monitoring using wearables for cardiovascular risk assessment: a systematic review.
- Method-comparison study between a watch-like sensor and a cuff-based device for 24-h ambulatory blood pressure monitoring. [conflict statement: the authors' institute holds the patent on the method tested, and two authors are its inventors]
- Clinical Validation of Photoplethysmography-Based Finger Cuffless Blood Pressure Monitoring Following European Society of Hypertension Recommendations.