Respiratory Rate Research Desk
The working file on Respiratory Rate. Each claim below is rated on its own evidence.
respiratory rate hub
Evidence files:
HRV·RHR·Sleep·VO2 Max·Wearable ECG·CGM·Recovery·Activity·Blood Oxygen·Stress·Temperature
Do consumer wearables and contactless sensors measure respiratory rate accurately?
§
Consumer wearable and contactless devices can measure respiratory rate accurately enough to be meaningful for everyday users.
- Systematic-review or meta-analytic evidence affirms the core relationship.
- Multiple supporting studies are consistent with the core relationship.
- An anchor study is present but its own reported finding does not cleanly affirm this claim, so it does not count toward the Established threshold.
- A clean anchor supports this claim, but an unresolved conflicting study contradicts it on the core relationship, so the evidence is treated as in genuine dispute.
- Some evidence conflicts on secondary points that do not change the core direction.
48 studies scored · 3 anchor · 42 supporting · 3 conflicting · Last reviewed Aug 1, 2026
Studies in Pool48
Claims Rated1
Last PubMed SyncAug 1, 2026
Next Scheduled ReviewAug 8, 2026
3 systematic reviews
37 observational
4 validity studies
4 other
MAINTAINED BY
The KnowYourPrime Research Desk · reviewed against new evidence on a rolling basis
Evidence for: Consumer wearable and contactless devices can measure respiratory rate accurately enough to be meaningful for everyday users.
48 scored
Anchor Studies3
Novel wearable and contactless heart rate, respiratory rate, and oxygen saturation monitoring devices: a systematic review and meta-analysis.
Journal
Anaesthesia · 2022
PMID
35947876
RCR
3.7
Citations
29
Study type
Meta-analysis
Sample size
[ not stated in the abstract ]
heart ratehumans
Effectiveness of consumer-grade contactless vital signs monitors: a systematic review and meta-analysis.
Journal
Journal of clinical monitoring and computing · 2022
PMID
34240262
RCR
2.9
Citations
21
Study type
Meta-analysis
Sample size
[ not stated in the abstract ]
adultheart rate
Clinical validation of a wearable respiratory rate device: A brief report.
Journal
Chronic respiratory disease · 2023
PMID
37612250
RCR
1.4
Citations
6
Study type
Observational study
Sample size
[ not stated in the abstract ]
humansbody mass index
Supporting Studies42
Clinical evaluation of a smart wristband for monitoring oxygen saturation, pulse rate, and respiratory rate.
Journal
Journal of clinical monitoring and computing · 2025
PMID
39388061
RCR
4.8
Citations
10
Study type
Observational study
Sample size
[ not stated in the abstract ]
adultaged
Reliable Contactless Monitoring of Heart Rate, Breathing Rate, and Breathing Disturbance During Sleep in Aging: Digital Health Technology Evaluation Study.
Journal
JMIR mHealth and uHealth · 2024
PMID
39190477
RCR
3.0
Citations
11
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansaged
Reliability of Zephyr BioHarness Respiratory Rate at Rest, During the Modified Canadian Aerobic Fitness Test and Recovery.
Journal
Journal of strength and conditioning research · 2020
PMID
30694966
RCR
1.0
Citations
10
Study type
Primary study
Sample size
[ not stated in the abstract ]
adultexercise
Validating a Consumer Smartwatch for Nocturnal Respiratory Rate Measurements in Sleep Monitoring.
Journal
Sensors (Basel, Switzerland) · 2026
PMID
37766031
RCR
0.6
Citations
4
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansmale
Agreement between standard and continuous wireless vital sign measurements after major abdominal surgery: a clinical comparison study.
Journal
Physiological measurement · 2022
PMID
36322987
RCR
1.4
Citations
9
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansmonitoring, physiologic
Respiratory rate assessments using a dual-accelerometer device.
Journal
Respiratory physiology & neurobiology · 2014
PMID
24263211
RCR
1.7
Citations
30
Study type
Primary study
Sample size
[ not stated in the abstract ]
accelerometryadult
Monitoring respiratory rates with a wearable system using a stretchable strain sensor during moderate exercise.
Journal
Medical & biological engineering & computing · 2020
PMID
31734768
RCR
1.1
Citations
22
Study type
Primary study
Sample size
[ not stated in the abstract ]
adultexercise
Validation of Instantaneous Respiratory Rate Using Reflectance PPG from Different Body Positions.
Journal
Sensors (Basel, Switzerland) · 2019
PMID
30384462
RCR
1.1
Citations
14
Study type
Validity study
Sample size
[ not stated in the abstract ]
adultalgorithms
Morphic Sensors for Respiratory Parameters Estimation: Validation against Overnight Polysomnography.
Journal
Biosensors · 2023
PMID
37504102
RCR
0.5
Citations
3
Study type
Primary study
Sample size
[ not stated in the abstract ]
malefemale
Clinical validation of a novel accelerometer-based respiratory rate monitor.
Journal
Journal of clinical monitoring and computing · 2026
PMID
41649789
Citations
0
Study type
Validity study
Sample size
[ not stated in the abstract ]
humansrespiratory rate
Bluetooth low energy thermohygrometric sensor for continuous respiratory rate monitoring: 48 h breathing simulator and human proof-of-concept.
Journal
Biomedical physics & engineering express · 2026
PMID
41880662
Citations
0
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansrespiratory rate
Seconds Matter: Rapid Non-Contact Monitoring of Heart and Respiratory Rate from Face Videos.
Journal
Sensors (Basel, Switzerland) · 2026
PMID
41829469
Citations
0
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansheart rate
Feasibility of Monitoring Heart and Respiratory Rates Using Nonwearable Devices and Consistency of the Measured Parameters: Pilot Feasibility Study.
Journal
JMIR human factors · 2024
PMID
39378444
RCR
0.5
Citations
1
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansfeasibility studies
A Guide to Measuring Heart and Respiratory Rates Based on Off-the-Shelf Photoplethysmographic Hardware and Open-Source Software.
Journal
Sensors (Basel, Switzerland) · 2024
PMID
38931550
RCR
0.6
Citations
2
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansphotoplethysmography
Evaluation of Photoplethysmography-Based Monitoring of Respiration Rate During High-Intensity Interval Training: Implications for Healthcare Monitoring.
Journal
Biosensors · 2024
PMID
39727896
RCR
0.5
Citations
1
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansphotoplethysmography
Accuracy of the Instantaneous Breathing and Heart Rates Estimated by Smartphone Inertial Units.
Journal
Sensors (Basel, Switzerland) · 2025
PMID
40006324
RCR
2.7
Citations
6
Study type
Primary study
Sample size
[ not stated in the abstract ]
humanssmartphone
Validity of a novel respiratory rate monitor comprising stretchable strain sensors during a 6-min walking test in patients with chronic pulmonary obstructive disease.
Journal
Respiratory medicine · 2022
PMID
34768076
RCR
0.2
Citations
2
Study type
Validity study
Sample size
[ not stated in the abstract ]
adultaged
Accuracy of Rhythm Diagnostic Systems' MultiSense ® in Detection of Arterial Oxygen Saturation and Respiratory Rate During Hypoxia in Humans: Effects of Skin Color and Device Localization.
Journal
Sensors (Basel, Switzerland) · 2025
PMID
39796917
RCR
0.0
Citations
0
Study type
Primary study
Sample size
[ not stated in the abstract ]
adulthumans
Prospective observational study of 2 wearable strain sensors for measuring the respiratory rate.
Journal
Medicine · 2024
PMID
39029069
RCR
0.0
Citations
0
Study type
Observational study
Sample size
[ not stated in the abstract ]
humansrespiratory rate
Contact-Based Methods for Measuring Respiratory Rate.
Journal
Sensors (Basel, Switzerland) · 2019
PMID
30795595
RCR
11.9
Citations
148
Study type
Review
Sample size
[ not stated in the abstract ]
exercisehumans
Breathing Rate Estimation From the Electrocardiogram and Photoplethysmogram: A Review.
Journal
IEEE reviews in biomedical engineering · 2018
PMID
29990026
RCR
10.5
Citations
158
Study type
Review
Sample size
[ not stated in the abstract ]
algorithmselectrocardiography
An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogram.
Journal
Physiological measurement · 2016
PMID
27027672
RCR
8.9
Citations
150
Study type
Primary study
Sample size
[ not stated in the abstract ]
adolescentadult
Toward a Robust Estimation of Respiratory Rate From Pulse Oximeters.
Journal
IEEE transactions on bio-medical engineering · 2017
PMID
27875128
RCR
7.2
Citations
105
Study type
Validity study
Sample size
[ not stated in the abstract ]
adolescentadult
Photoplethysmographic derivation of respiratory rate: a review of relevant physiology.
Journal
Journal of medical engineering & technology · 2012
PMID
22185462
RCR
5.0
Citations
102
Study type
Review
Sample size
[ not stated in the abstract ]
humansmonitoring, physiologic
Comparison of different modulations of photoplethysmography in extracting respiratory rate: from a physiological perspective.
Journal
Physiological measurement · 2021
PMID
32731213
RCR
2.8
Citations
28
Study type
Comparative study
Sample size
[ not stated in the abstract ]
algorithmsarm
Non-Contact Respiratory Monitoring Using an RGB Camera for Real-World Applications.
Journal
Sensors (Basel, Switzerland) · 2021
PMID
34372363
RCR
2.7
Citations
19
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansmonitoring, physiologic
An IMU-Based Wearable System for Respiratory Rate Estimation in Static and Dynamic Conditions.
Journal
Cardiovascular engineering and technology · 2023
PMID
36849621
RCR
2.4
Citations
9
Study type
Primary study
Sample size
[ not stated in the abstract ]
malehumans
Contactless Camera-Based Heart Rate and Respiratory Rate Monitoring Using AI on Hardware.
Journal
Sensors (Basel, Switzerland) · 2023
PMID
37177754
RCR
2.2
Citations
6
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansheart rate
Respiratory Rate Estimation during Walking and Running Using Breathing Sounds Recorded with a Microphone.
Journal
Biosensors · 2023
PMID
37367002
RCR
1.7
Citations
9
Study type
Primary study
Sample size
[ not stated in the abstract ]
malefemale
Electrocardiogram Derived Respiratory Rate Using a Wearable Armband.
Journal
IEEE transactions on bio-medical engineering · 2021
PMID
32746038
RCR
1.4
Citations
12
Study type
Primary study
Sample size
[ not stated in the abstract ]
algorithmselectrocardiography
Comparative evaluation of susceptibility to motion artifact in different wearable systems for monitoring respiratory rate.
Journal
IEEE transactions on information technology in biomedicine : a publication of the IEEE Engineering in Medicine and Biology Society · 2010
PMID
20007035
RCR
1.3
Citations
32
Study type
Comparative study
Sample size
[ not stated in the abstract ]
adultartifacts
Estimation of Heart Rate and Respiratory Rate from PPG Signal Using Complementary Ensemble Empirical Mode Decomposition with both Independent Component Analysis and Non-Negative Matrix Factorization.
Journal
Sensors (Basel, Switzerland) · 2021
PMID
32517226
RCR
1.3
Citations
13
Sample size
[ not stated in the abstract ]
algorithmsheart rate
A novel computational signal processing framework towards multimodal vital signs extraction using neck-worn wearable devices.
Journal
Scientific reports · 2024
PMID
39333140
RCR
1.3
Citations
2
Study type
Primary study
Sample size
[ not stated in the abstract ]
humanswearable electronic devices
Finger and forehead PPG signal comparison for respiratory rate estimation.
Journal
Physiological measurement · 2020
PMID
31422948
RCR
1.2
Citations
15
Study type
Primary study
Sample size
[ not stated in the abstract ]
adultfemale
Feasibility of Heart Rate and Respiratory Rate Estimation by Inertial Sensors Embedded in a Virtual Reality Headset.
Journal
Sensors (Basel, Switzerland) · 2021
PMID
33327531
RCR
0.7
Citations
8
Study type
Primary study
Sample size
[ not stated in the abstract ]
ballistocardiographyfeasibility studies
Respiratory modulation of the heart rate: A potential biomarker of cardiorespiratory function in human.
Journal
Computers in biology and medicine · 2024
PMID
38564855
RCR
0.5
Citations
1
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansmale
Respiratory Rate Monitoring via a Fibre Bragg Grating-Embedded Respirator Mask with a Wearable Miniature Interrogator.
Journal
Sensors (Basel, Switzerland) · 2024
PMID
39686013
Citations
1
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansrespiratory rate
A systematic review of contactless respiratory rate measurement using RGB cameras.
Journal
Physiological measurement · 2025
PMID
40816317
Citations
2
Study type
Systematic review
Sample size
[ not stated in the abstract ]
respiratory ratehumans
Comparison of Techniques for Respiratory Rate Extraction from Electrocardiogram and Photoplethysmogram.
Journal
Sensors (Basel, Switzerland) · 2025
PMID
40871998
Citations
2
Study type
Comparative study
Sample size
[ not stated in the abstract ]
photoplethysmographyhumans
Continuous respiratory rate monitoring through temporal fusion of ECG and PPG signals.
Journal
PloS one · 2025
PMID
40526643
Citations
2
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansrespiratory rate
Toward accurate estimation of respiratory rate from the photoplethysmogram: effect of different window period of PPG signals.
Journal
Physical and engineering sciences in medicine · 2025
PMID
40586998
Citations
1
Study type
Primary study
Sample size
[ not stated in the abstract ]
photoplethysmographyhumans
Accelerometer-based estimation of respiratory rate using principal component analysis and autocorrelation.
Journal
Physiological measurement · 2025
PMID
40054067
Citations
1
Study type
Primary study
Sample size
[ not stated in the abstract ]
humansaccelerometry
Conflicting Studies3
Can Caregivers Reliably Assess Their Child's Heart Rate and Respiration Rate Using Smartphone or Smartwatch Applications?
Journal
Pediatric emergency care · 2025
PMID
40341066
Citations
0
Study type
Primary study
Sample size
[ not stated in the abstract ]
Bears on
Core claim
humanschild
Comparison of manual and automated respiratory rate measurements on hospital wards: a prospective observational study.
Journal
Journal of clinical monitoring and computing · 2026
PMID
41240250
Citations
0
Study type
Comparative study
Sample size
[ not stated in the abstract ]
Bears on
Core claim
humansrespiratory rate
The use of a smart-textile garment during high-intensity functional training: a pilot study.
Journal
The Journal of sports medicine and physical fitness · 2019
PMID
30024125
RCR
0.6
Citations
13
Study type
Primary study
Sample size
[ not stated in the abstract ]
Bears on
Secondary point
adultclothing
01On Our Desk
In Progress
What we're reading right now
A study only enters a claim's evidence once it's read in full and weighed against what's already in file.
Identified
Queued
Under Review
Extracted
Pending Consensus
Nothing on the desk this cycle.
02Research Log
AUG 1 2026
doi:10.1007/s10877-021-00734-9 entered as anchor on the measurement_validity claim.
doi:10.1007/s10877-024-01229-z entered as supporting on the measurement_validity claim.
doi:10.1007/s10877-025-01380-1 entered as conflicting on the measurement_validity claim.
doi:10.1007/s10877-026-01419-x entered as supporting on the measurement_validity claim.
doi:10.1007/s11517-019-02062-2 entered as supporting on the measurement_validity claim.
doi:10.1007/s13239-023-00657-3 entered as supporting on the measurement_validity claim.
doi:10.1007/s13246-025-01582-6 entered as supporting on the measurement_validity claim.
doi:10.1016/j.compbiomed.2024.108335 entered as supporting on the measurement_validity claim.
doi:10.1016/j.resp.2013.11.003 entered as supporting on the measurement_validity claim.
doi:10.1016/j.rmed.2021.106675 entered as supporting on the measurement_validity claim.
doi:10.1038/s41598-024-72184-7 entered as supporting on the measurement_validity claim.
doi:10.1088/0967-3334/37/4/610 entered as supporting on the measurement_validity claim.
doi:10.1088/1361-6579/ab3be0 entered as supporting on the measurement_validity claim.
doi:10.1088/1361-6579/abaaf0 entered as supporting on the measurement_validity claim.
doi:10.1088/1361-6579/ac9fa3 entered as supporting on the measurement_validity claim.
doi:10.1088/1361-6579/adbe23 entered as supporting on the measurement_validity claim.
doi:10.1088/1361-6579/adfc24 entered as supporting on the measurement_validity claim.
doi:10.1088/2057-1976/ae5710 entered as supporting on the measurement_validity claim.
doi:10.1097/md.0000000000038818 entered as supporting on the measurement_validity claim.
doi:10.1097/pec.0000000000003400 entered as conflicting on the measurement_validity claim.
doi:10.1109/rbme.2017.2763681 entered as supporting on the measurement_validity claim.
doi:10.1109/tbme.2016.2613124 entered as supporting on the measurement_validity claim.
doi:10.1109/tbme.2020.3004730 entered as supporting on the measurement_validity claim.
doi:10.1109/titb.2009.2037614 entered as supporting on the measurement_validity claim.
doi:10.1111/anae.15834 entered as anchor on the measurement_validity claim.
doi:10.1177/14799731231198865 entered as anchor on the measurement_validity claim.
doi:10.1371/journal.pone.0325307 entered as supporting on the measurement_validity claim.
doi:10.1519/jsc.0000000000003046 entered as supporting on the measurement_validity claim.
doi:10.2196/53643 entered as supporting on the measurement_validity claim.
doi:10.2196/56547 entered as supporting on the measurement_validity claim.
doi:10.23736/s0022-4707.18.08689-9 entered as conflicting on the measurement_validity claim.
doi:10.3109/03091902.2011.638965 entered as supporting on the measurement_validity claim.
doi:10.3390/bios13060637 entered as supporting on the measurement_validity claim.
doi:10.3390/bios13070703 entered as supporting on the measurement_validity claim.
doi:10.3390/bios14120631 entered as supporting on the measurement_validity claim.
doi:10.3390/s18113705 entered as supporting on the measurement_validity claim.
doi:10.3390/s19040908 entered as supporting on the measurement_validity claim.
doi:10.3390/s20113238 entered as supporting on the measurement_validity claim.
doi:10.3390/s20247168 entered as supporting on the measurement_validity claim.
doi:10.3390/s21155126 entered as supporting on the measurement_validity claim.
doi:10.3390/s23094550 entered as supporting on the measurement_validity claim.
doi:10.3390/s23187976 entered as supporting on the measurement_validity claim.
doi:10.3390/s24123766 entered as supporting on the measurement_validity claim.
doi:10.3390/s24237476 entered as supporting on the measurement_validity claim.
doi:10.3390/s25010127 entered as supporting on the measurement_validity claim.
doi:10.3390/s25041094 entered as supporting on the measurement_validity claim.
doi:10.3390/s25165136 entered as supporting on the measurement_validity claim.
doi:10.3390/s26051506 entered as supporting on the measurement_validity claim.
AUG 1 2026
doi:10.1007/s10877-025-01380-1 entered as a conflicting study on the measurement_validity claim, bearing on the core claim.
It caps the rating until reconciled.
AUG 1 2026
doi:10.1097/pec.0000000000003400 entered as a conflicting study on the measurement_validity claim, bearing on the core claim.
It caps the rating until reconciled.
AUG 1 2026
doi:10.23736/s0022-4707.18.08689-9 entered as a conflicting study on the measurement_validity claim, flagged as bearing on a secondary point, not the core direction.
Because it does not hit the core claim, it does not cap the rating and is tracked as an open question.
04Revision Log
AUG 1 2026
measurement_validity: opened at Uncertain
Opened at Uncertain by rule from the current evidence for: Consumer-grade wearable or contactless devices measure respiratory rate with accuracy comparable to clinical reference methods..