Pulse Oximeter Accuracy and Skin Tone

Question assessedDoes skin pigmentation change how accurately a pulse oximeter, including one in a wearable, reads blood oxygen?

KnowYourPrime evidence ratingUncertainEvidence is conflicting, inconsistent, or difficult to interpret.

The largest and most recent studies find oximeters read blood oxygen higher in people with darker skin, which can hide low oxygen, but several careful studies found no difference and one meta-analysis judged the gap may not be clinically meaningful. The direction is consistent where a difference is found; its size and importance are disputed.

Finding KYP-ER-0002Version 1.0Reviewed Rating rule 1.4CC BY 4.0

Why this is rated Uncertain

An oximeter that reads high is the dangerous direction, because it can show a reassuring number while arterial oxygen is low. The claim is that this error varies with skin pigmentation.

A meta-analysis of 23 oximetry studies and a large intensive care study with spectrophotometer-measured skin tone both found higher readings for darker skin, and controlled desaturation trials going back to 2005 found the same. Against that, two emergency and intensive care studies with objective or standardized pigmentation measures found no significant difference, and a recent ICU study found no association between skin tone and oximeter error. The rule rates the claim Uncertain because affirming and contradicting evidence both bear on the core question.

Almost all of this evidence comes from finger and clinical oximeters. Wearables use reflected light at the wrist, which has been tested far less.

How the rating rule reads this evidence
  • Systematic-review or meta-analytic evidence affirms the core relationship.
  • Multiple supporting studies are consistent with the core relationship.
  • An anchor study is present but its own reported finding does not cleanly affirm this claim, so it does not count toward the Established threshold.
  • A clean anchor supports this claim, but an unresolved conflicting study contradicts it on the core relationship, so the evidence is treated as in genuine dispute.
  • Null-result evidence bears directly on the claimed effect.

Key results

What this finding covers

Applies to: Clinical and fingertip pulse oximeters compared with arterial blood, across skin tone groups, plus wearable pulse-rate accuracy by skin tone.

Does not cover: Direct tests of wrist-worn smartwatch blood oxygen across skin tones, which are scarce; any single device brand; medical decisions about oxygen therapy.

Limitations

Evidence base

23 studies scored · 7 anchor · 1 conflicting · 1 null · 14 supporting · 13 letters or comments not counted · Evidence last evaluated Oct 25, 2026

Anchor studies

The studies the rating rests on.

  1. The impact of skin tone on performance of pulse oximeters used by NHS England COVID Oximetry @home scheme: measurement and diagnostic accuracy study.Martin et al., BMJ, 2026. doi.org/10.1136/bmj-2025-085535
  2. Do Differences in Skin Pigmentation Affect Detection of Hypoxemia by Pulse Oximetry: A Systematic Review of the Literature.Cotton et al., Clinical Nursing Research, 2025. doi.org/10.1177/10547738251374746
  3. Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis.Singh et al., Journal of Medical Internet Research, 2024. doi.org/10.2196/62769 [author affiliations include Verily Life Sciences]
  4. Low Perfusion and Missed Diagnosis of Hypoxemia by Pulse Oximetry in Darkly Pigmented Skin: A Prospective Study.Gudelunas et al., Anesthesia and Analgesia, 2024. doi.org/10.1213/ANE.0000000000006755
  5. Effects of skin pigmentation on pulse oximeter accuracy at low saturation.Bickler et al., Anesthesiology, 2005. doi.org/10.1097/00000542-200504000-00004
  6. Effect of skin pigmentation on pulse oximetry accuracy in the emergency department.Adler et al., Academic Emergency Medicine, 1998. doi.org/10.1111/j.1553-2712.1998.tb02772.x
  7. Accuracy of pulse oximetry in pigmented patients.Bothma et al., South African Medical Journal, 1997. pubmed.ncbi.nlm.nih.gov/8914569/

Conflicting studies

Studies whose findings run against the claim.

  1. Accuracy and Bias of Pulse Oximetry in the Intensive Care Unit: A Prospective Observational Study.Barros et al., Nursing in Critical Care, 2026. doi.org/10.1111/nicc.70433

Null results

Studies that tested the claim and found no effect.

  1. Effect of Skin Pigmentation and Finger Choice on Accuracy of Oxygen Saturation Measurement in an IoT-Based Pulse Oximeter.Haxha et al., Sensors, 2024. doi.org/10.3390/s24113301

Supporting studies

Studies consistent with the claim that do not carry the rating on their own.

Show all 14 supporting studies
  1. Broad-linewidth sources result in a skin-tone bias in noninvasive optical measurement of oxygen saturation.Blaney et al., Journal of Biomedical Optics, 2026. doi.org/10.1117/1.JBO.31.7.070501
  2. Pilot Study: Using Reflectance as a Proxy for Skin Pigmentation to Assess Pulse Oximeter Accuracy.Branstad et al., Hospital Pediatrics, 2026. doi.org/10.1542/hpeds.2025-008471
  3. Monte Carlo Simulation of the Effect of Melanin Concentration on Light-Tissue Interactions in Reflectance Pulse Oximetry.Al-Halawani et al., Sensors, 2025. doi.org/10.3390/s25020559
  4. Pulse oximeter bench tests under different simulated skin tones.Swamy et al., Medical & Biological Engineering & Computing, 2025. doi.org/10.1007/s11517-024-03091-2
  5. Determining factors of pulse oximetry accuracy: a literature review.León-Valladares et al., Revista Clinica Espanola, 2024. doi.org/10.1016/j.rceng.2024.04.005
  6. Dual-ratio approach to pulse oximetry and the effect of skin tone.Blaney et al., Journal of Biomedical Optics, 2024. doi.org/10.1117/1.JBO.29.S3.S33311
  7. Exploring the bias: how skin color influences oxygen saturation readings via Monte Carlo simulations.Swamy et al., Journal of Biomedical Optics, 2024. doi.org/10.1117/1.JBO.29.S3.S33308
  8. Monte Carlo simulation of the effect of melanin concentration on light-tissue interactions for transmittance pulse oximetry measurement.Al-Halawani et al., Journal of Biomedical Optics, 2024. doi.org/10.1117/1.JBO.29.S3.S33305
  9. Pulse Oximetry: 2023 Year in Review.Hess et al., Respiratory Care, 2024. doi.org/10.4187/respcare.12023
  10. Pulse Oximeter Performance, Racial Inequity, and the Work Ahead.Okunlola et al., Respiratory Care, 2022. doi.org/10.4187/respcare.09795
  11. Racial Disparities in Occult Hypoxemia and Clinically Based Mitigation Strategies to Apply in Advance of Technological Advancements.Chesley et al., Respiratory Care, 2022. doi.org/10.4187/respcare.09769
  12. Racial Disparity in Oxygen Saturation Measurements by Pulse Oximetry: Evidence and Implications.Jamali et al., Annals of the American Thoracic Society, 2022. doi.org/10.1513/AnnalsATS.202203-270CME
  13. Self-reported Race/Ethnicity and Intraoperative Occult Hypoxemia: A Retrospective Cohort Study.Burnett et al., Anesthesiology, 2022. doi.org/10.1097/ALN.0000000000004153
  14. Potential errors in pulse oximetry. I. Pulse oximeter evaluation.Ralston et al., Anaesthesia, 1991. doi.org/10.1111/j.1365-2044.1991.tb09410.x

Version history

VersionDateRatingChange
1.0October 24, 2026UncertainFirst published.

Rating history on the research desk: Uncertain (July 21, 2026).

Cite this finding

KnowYourPrime Evidence Registry, Finding KYP-ER-0002, version 1.0, reviewed 2026-10-24. https://knowyourprime.com/evidence-pulse-oximeter-skin-tone-bias

Machine-readable: JSON · Markdown · All findings

Related

This is KnowYourPrime's own assessment of the published evidence, rated under the KnowYourPrime evidence methodology. The rating is set by a fixed rule from the roles of the studies above and moves only when the evidence or the rule changes. Every change to the rating or the text is published as a new dated version. To report an error, write to corrections@knowyourprime.com.